High Potential Surgical Generator Testing
Magnelab’s high voltage isolation transformer 3304-xxx Series is designed in cooperation with Underwriters Laboratory (UL) to be used for high potential (Hi-Pot) testing of surgical generators.
Customized Sizes and Ratios with a Primary Focus on Patient Safety
Several turns ratios are available to accommodate input and output requirements. Custom ratios and other parameters may be available upon request. This unit fits a standard banana plug. Each transformer has 10,000 volt isolation for exceptional patient safety. Dimension sizes vary from 0.300 inches to 2.50 inches (7.61 mm to 63.49 mm). Turn ratios can go from 1:1.1 to 1:10 (see turn ratios specifications below). This product comes with a standard 10-year warranty.
MAGNELAB Part Number |
Turns Ratio | Primary Turns | Secondary Turns | MAXIMUM PRIMARY VOLTAGE
PEAK TO PEAK |
|||
100 KHz | 250 KHz | 500 KHz |
750 KHz |
||||
3304-005 |
1:4 | 15 | 60 | 460V | 1,150V | 2,300V | 3,450V |
3304-006 | 1:2 | 15 | 30 | 460V | 1,150V | 2,300V |
3,450V |
3304-008 |
1:1.5 | 30 | 45 | 920V | 2,300V | 4,600V | 6,900V |
3304-009 | 1:1.1 | 30 | 33 | 460V | 2,300V | 4,600V |
6,900V |
3304-010 |
1:7 | 15 | 105 | 460V | 1,150V | 2,300V | 3,450V |
3304-012 | 1:8 | 15 | 120 | 460V | 1,150V | 2,300V |
3,450V |
3304-013 |
1:10 | 15 | 150 | 460V | 1,150V | 2,300V | 3,450V |
3304-014 | 1:3 | 30 | 90 | 920V | 2,300V | 4,600V |
6,900V |